| FEATURES |
| Formation conductivity measurement
in wet/dry boreholes or through plastic casing |
| Focussed measurements almost eliminates
borehole signal |
| PSD (phase-sensitive detector) discriminates
between magnetic susceptibility and conductivity signals |
| Highly stable measurement is almost
independent of temperature |
| Option of natural gamma for correlation |
| MEASUREMENTS |
| Formation conductivity |
| Natural gamma (optional) |
| APPLICATIONS |
| Water |
Indicator of permeable zones and
porosity
Formation-water salinity
Long-term well monitoring |
| Engineering/Minerals |
Ore identification and quality
Correlation |
| Other |
Indication of hydrocarbons |
| OPERATING
CONDITIONS |
| Borehole type |
open/ plastic or grp cased, air/water
filled |
| Centralisation |
none; fin standoff recommended |
| SPECIFICATIONS |
| Diameter |
38mm |
| Length |
2.15m |
| Weight |
9kg |
| Max. temperature |
70ºC (extended range available) |
| Max. pressure |
20MPa (extended range available) |
| Number of coils |
4 |
| Operating frequency |
39kHz |
| Effective TX-RX spacing |
500mm (20in) |
| Resolution |
1mS/m |
| Accuracy |
3% of reading over whole temperature
range |
| Conductivity range |
2 to 3000 mS/m |
| Natural-gamma detector |
50mm x 25mm NaI(Tl) scintillation
crystal (larger sizes available) |
| SALES
INFORMATION |
| Probe |
| 25 060 000 |
Focussed Induction probe with calibration
ring |
| 25 061 000 |
includes natural gamma |
| Accessories |
| 23 001 000 |
Fin stand-off |
| 20 070 000 |
Natural-gamma API calibrator without
source |
| 30 010 000 |
3.7MBq 137Cs source for natural
gamma calibrator |